• DocumentCode
    3148073
  • Title

    Noninvasive blood pressure analysis using empirical mode decomposition (EMD) - Based oscillometric method

  • Author

    Hsiao, Fu-Jung ; Hung, Chen-Chuan

  • Author_Institution
    Center for Meas. Stand., Ind. Technol. Res. Inst., Chutung, Taiwan
  • Volume
    3
  • fYear
    2010
  • fDate
    16-18 Oct. 2010
  • Firstpage
    973
  • Lastpage
    977
  • Abstract
    This study proposes a new method to extract oscillometric pulses for noninvasive blood pressure measurement using empirical mode decomposition, based on the oscillometric method. The mean and standard deviation of the present method in systolic and diastolic pressure are 1.6 and 5.2, and 0.7 and 5.2, respectively, which not only passes the AAMI (Association for the Advancement of Medical Instrumentation) and EN (European Norm) blood pressure measurement criteria, but also performs more accurate measurements than commercial noninvasive blood pressure measurement devices.
  • Keywords
    biomedical equipment; blood pressure measurement; diastolic pressure; empirical mode decomposition; noninvasive blood pressure measurement devices; oscillometric method; standard deviation; systolic pressure; Accuracy; Biomedical monitoring; Blood pressure; Distortion measurement; Pressure measurement; Pulse measurements; empirical mode decomposition; noninvasive blood pressure measurement; oscillometric method;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2010 3rd International Conference on
  • Conference_Location
    Yantai
  • Print_ISBN
    978-1-4244-6495-1
  • Type

    conf

  • DOI
    10.1109/BMEI.2010.5639803
  • Filename
    5639803