  • Abstract
    Due to international standards each implantable cardiac pacemaker reacts to the application of a permanent magnet by changing the pacing rate in a predefined way depending on the depletion level of its battery. The purpose of this study has been to use a routine ECG recording to verify in a telemedicine framework whether the pacemaker works correctly or if further examination is indicated. The system has been implemented prototypically and has been evaluated in a clinical pilot trial on 24 consecutive patients with a total of 17 different pacemaker models from 6 different manufacturers. The promising results indicate that the presented follow-up concept, which can be handled by general practitioners, has the potential to work as an efficient screening method and may spare a significant number of patients the burden of having to travel to specialized hospitals which are often far away from patients´ homes
  • Keywords
    biomedical equipment; electrocardiography; pacemakers; permanent magnets; telemedicine; ECG recording; battery; implantable cardiac pacemaker; international standards; manufacturer independent remote pacemaker follow-up; permanent magnet; screening method; telemedicine; Batteries; Cardiology; Collaboration; Electrocardiography; Guidelines; Manufacturing; Medical services; Pacemakers; Permanent magnets; Telemedicine;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2005
  • Conference_Location
    Lyon
  • Print_ISBN
    0-7803-9337-6
  • Type

    conf

  • DOI
    10.1109/CIC.2005.1588030
  • Filename
    1588030