• DocumentCode
    3174141
  • Title

    Quantitative coronary arteriography: improving catheter calibration by an efficient correction for differential magnification

  • Author

    Fischer, F. ; Wunderlich, W. ; Röhrig, B. ; Kristiansen, G. ; Linderer, T. ; Schultheiss, HP

  • Author_Institution
    Dept. of Cardiology, Freie Univ. Berlin, Germany
  • fYear
    1995
  • fDate
    10-13 Sept. 1995
  • Firstpage
    553
  • Lastpage
    556
  • Abstract
    Differential magnification of the vessel and of the scaling catheter in the X-ray field causes significant errors in calibration of imaged vessel size. To correct for differential magnification properly, laborious and costly calculations are required. The authors derived the equations of accurate correction for differential magnification. Starting from these equations the authors reduced the correction effort to the acquisition of a second complementary angiogram and two digitizing operations. The validation of this approximate procedure in computer simulations and in vitro phantom studies showed a correction for differential magnification up to 85%.
  • Keywords
    angiocardiography; calibration; diagnostic radiography; blood vessel; catheter calibration improvement; computer simulations; differential magnification; digitizing operations; efficient correction; imaged vessel size; in vitro phantom studies; medical diagnostic imaging; quantitative coronary arteriography; second complementary angiogram; Angiography; Calibration; Cardiology; Catheters; Computer simulation; Delta modulation; Equations; Error correction; Imaging phantoms; In vitro;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology 1995
  • Conference_Location
    Vienna, Austria
  • Print_ISBN
    0-7803-3053-6
  • Type

    conf

  • DOI
    10.1109/CIC.1995.482724
  • Filename
    482724