• DocumentCode
    3179279
  • Title

    Accurate phase measurement with two sampling voltmeters

  • Author

    Lapuh, Rado

  • Volume
    1
  • fYear
    2001
  • fDate
    21-23 May 2001
  • Firstpage
    645
  • Abstract
    An accurate phase measurement technique is described. The technique uses two sampling voltmeters for signals with voltage ratio up to 1:1000 and a single sampling voltmeter with input switch for signals with voltage ratio up to 1:10. At 50 Hz, uncertainties below 0.05 m° were attained at voltages below 8 V and voltage ratios up to 1:10. This uncertainty rose to 15 m° at 10 kHz. When corrected for range-to-range phase differences, only slightly higher uncertainties are attainable for voltage ratios up to 1:1000. A technique for the automatic correction of range-to-range phase differences is described
  • Keywords
    digital multimeters; digital voltmeters; measurement uncertainty; phase measurement; signal sampling; 10 kHz; 50 Hz; DMM; accurate phase measurement technique; automatic correction; digital measurement; input switch; linearity; range switching effect; range-to-range phase differences; repeatability; sine fitting; two sampling voltmeters; uncertainties; Bridge circuits; Frequency; Goniometers; Jitter; Linearity; Phase measurement; Sampling methods; Switches; Voltage; Voltmeters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2001. IMTC 2001. Proceedings of the 18th IEEE
  • Conference_Location
    Budapest
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-6646-8
  • Type

    conf

  • DOI
    10.1109/IMTC.2001.928897
  • Filename