DocumentCode
3191603
Title
Removing cardiac interference from the electroencephalogram using a modified Pan-Tompkins algorithm and linear regression
Author
Waser, M. ; Garn, H.
Author_Institution
AIT Austrian Inst. of Technol. GmbH, Vienna, Austria
fYear
2013
fDate
3-7 July 2013
Firstpage
2028
Lastpage
2031
Abstract
Cardiac interference can alter the results of quantitative electroencephalograms (qEEG) used for medical diagnoses. The methods currently employed for the automated removal of cardiac interference, which rely solely on the electroencephalogram (EEG), are susceptible to non-cardiac interference commonly encountered in EEGs. Methods that rely on the electrocardiogram (ECG) - besides being unreliable when non-cardiac artifacts corrupt the ECG - either assume periodicity of the cardiac (QRS) peaks or alter uncorrupted EEG segments. This paper proposes a robust method for the automated removal of cardiac interference from EEGs by identifying QRS peaks in the ECG without assuming periodicity. Artificial signals consisting only of QRS peaks and the zero-lines in between are computed. Linear regression of the EEG channels on the “QRS signals” removes cardiac interference without altering uncorrupted EEG segments. The QRS-based regression method was tested on 30 multi-channel EEGs exhibiting cardiac interference of elderly subjects (15 male, 15 female). Achieving a correction rate of 80%, the QRS-based regression method has proved effective in removing cardiac interference from the EEG even in presence of additional non-cardiac interference in the EEG.
Keywords
electrocardiography; electroencephalography; geriatrics; interference (signal); medical signal processing; regression analysis; ECG noncardiac artifact; EEG channel linear regression; EEG noncardiac interference; QRS peak identification; QRS signal; QRS-based regression method; artificial signal; automated cardiac interference removal; cardiac peak periodicity; correction rate; elderly subject; electrocardiogram; medical diagnosis; modified Pan-Tompkins algorithm; multichannel EEG; qEEG; quantitative electroencephalogram; robust method; uncorrupted EEG segment alteration; zero-line; Electrocardiography; Electroencephalography; Filtering; Heart rate; Interference; Robustness; Shape;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location
Osaka
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2013.6609929
Filename
6609929
Link To Document