DocumentCode :
3193138
Title :
Effect of aluminium filtration on dose and image quality in paediatric slot-scanning radiography
Author :
Perks, T.D. ; Trauernicht, C. ; Hartley, Tom ; Hobson, C. ; Lawson, A. ; Scholtz, P. ; Dendere, Ronald ; Steiner, S. ; Douglas, T.S.
Author_Institution :
MRC/UCT Med. Imaging Res. Unit, Univ. of Cape T own, Observatory, South Africa
fYear :
2013
fDate :
3-7 July 2013
Firstpage :
2332
Lastpage :
2335
Abstract :
This paper examines the effect that a 1.8 mm aluminium filter has on paediatric patient dose and image quality for linear slot scanning radiography (LSSR). A dynamic dose prediction model for LSSR accurately predicted the dose reduction effects of added aluminium filtration. A cadaver imaging study was carried out to assess the effects of filtration on image quality. With 1.8 mm added aluminium filtration, no visible degradation to image contrast or clarity was found, and in some cases the aluminium filtration improved the image quality as judged by radiologists.
Keywords :
aluminium; diagnostic radiography; dosimetry; filtering theory; filtration; medical image processing; paediatrics; aluminium filter; aluminium filtration effect; cadaver imaging; dose quality; dynamic dose prediction model; image clarity; image contrast; image quality; linear slot scanning radiography; paediatric patient dose; paediatric slot-scanning radiography; size 1.8 mm; Aluminum; Cadaver; Filtration; Image quality; Pediatrics; Predictive models; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
ISSN :
1557-170X
Type :
conf
DOI :
10.1109/EMBC.2013.6610005
Filename :
6610005
Link To Document :
بازگشت