• DocumentCode
    3193138
  • Title

    Effect of aluminium filtration on dose and image quality in paediatric slot-scanning radiography

  • Author

    Perks, T.D. ; Trauernicht, C. ; Hartley, Tom ; Hobson, C. ; Lawson, A. ; Scholtz, P. ; Dendere, Ronald ; Steiner, S. ; Douglas, T.S.

  • Author_Institution
    MRC/UCT Med. Imaging Res. Unit, Univ. of Cape T own, Observatory, South Africa
  • fYear
    2013
  • fDate
    3-7 July 2013
  • Firstpage
    2332
  • Lastpage
    2335
  • Abstract
    This paper examines the effect that a 1.8 mm aluminium filter has on paediatric patient dose and image quality for linear slot scanning radiography (LSSR). A dynamic dose prediction model for LSSR accurately predicted the dose reduction effects of added aluminium filtration. A cadaver imaging study was carried out to assess the effects of filtration on image quality. With 1.8 mm added aluminium filtration, no visible degradation to image contrast or clarity was found, and in some cases the aluminium filtration improved the image quality as judged by radiologists.
  • Keywords
    aluminium; diagnostic radiography; dosimetry; filtering theory; filtration; medical image processing; paediatrics; aluminium filter; aluminium filtration effect; cadaver imaging; dose quality; dynamic dose prediction model; image clarity; image contrast; image quality; linear slot scanning radiography; paediatric patient dose; paediatric slot-scanning radiography; size 1.8 mm; Aluminum; Cadaver; Filtration; Image quality; Pediatrics; Predictive models; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
  • Conference_Location
    Osaka
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2013.6610005
  • Filename
    6610005