• DocumentCode
    3195612
  • Title

    Ultra-low-cost clinical pulse oximetry

  • Author

    Petersen, Christian Leth ; Heng Gan ; MacInnis, Martin J. ; Dumont, Guy A. ; Ansermino, J. Mark

  • Author_Institution
    Univ. of British Columbia, Vancouver, BC, Canada
  • fYear
    2013
  • fDate
    3-7 July 2013
  • Firstpage
    2874
  • Lastpage
    2877
  • Abstract
    An ultra-low-cost pulse oximeter is presented that interfaces a conventional clinical finger sensor with a mobile phone through the headset jack audio interface. All signal processing is performed using the audio subsystem of the phone. In a preliminary volunteer study in a hypoxia chamber, we compared the oxygen saturation obtained with the audio pulse oximeter against a commercially available (and FDA approved) reference pulse oximeter (Nonin Xpod). Good agreement was found between the outputs of the two devices.
  • Keywords
    headphones; mobile handsets; oximetry; signal processing equipment; Nonin Xpod; audio pulse oximeter; conventional clinical finger sensor; headset jack audio interface; hypoxia chamber; mobile phone; oxygen saturation; phone audio subsystem; reference pulse oximeter; signal processing; ultralow-cost clinical pulse oximetry; Blood; Diseases; Fingers; Light emitting diodes; Lungs; Mobile handsets; Pediatrics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
  • Conference_Location
    Osaka
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2013.6610140
  • Filename
    6610140