DocumentCode
320211
Title
Quantitative algorithm for the localization of the origin of ectopic ventricular beats using catheter pace mapping
Author
Latour, A. ; Grimbergen, C.A. ; Peeters, H.A.P. ; Linnenbank, A.C. ; SippensGroenewegen, A.
Author_Institution
Dept. of Med. Phys., Amsterdam Univ., Netherlands
Volume
3
fYear
1996
fDate
31 Oct-3 Nov 1996
Firstpage
1268
Abstract
In the treatment of a ventricular tachycardia by radiofrequency ablation, it is essential to position the tip of the catheter at the origin of the arrhythmia. A quantitative algorithm has been developed to estimate the site of origin of the arrhythmia on the basis of paced QRS integral maps and the corresponding spatial coordinates of the catheter tip positions. The algorithm has been validated by simulation, using catheter pace maps and spatial coordinates obtained in WPW-patients undergoing RF ablation. When assuming a coupled linear relation between coordinates of the catheter tip positions and the pace map, 71% of the required positions were estimated with an accuracy within the resolution of the position measurements
Keywords
electrocardiography; medical signal processing; RF ablation; WPW-patients; catheter pace maps; catheter tip positions; ectopic ventricular beats origin localization; paced QRS integral maps; position measurements resolution; quantitative algorithm; spatial coordinates; Catheters; Databases; Heart rate variability; Hospitals; Interpolation; Iterative algorithms; Medical treatment; Physics; Position measurement; Radio frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1996. Bridging Disciplines for Biomedicine. Proceedings of the 18th Annual International Conference of the IEEE
Conference_Location
Amsterdam
Print_ISBN
0-7803-3811-1
Type
conf
DOI
10.1109/IEMBS.1996.652805
Filename
652805