DocumentCode :
3211963
Title :
Characterization of an optical fiber thermometer with cladding mode suppression fiber
Author :
Yu, Qian ; Chen, Huacai
Author_Institution :
Coll. of Opt. & Electron. Technol., China Jiliang Univ., Hangzhou, China
Volume :
3
fYear :
2011
fDate :
29-31 July 2011
Abstract :
An optical fiber thermometer based on cladding mode suppression fiber was designed for medical applications in human body, e.g., hyperthermia treatment. In this paper, the basic sensing principle of fiber Bragg grating(FBG) was introduced, and the characteristic of FBG thermometer and the influence of bending curvature were investigated. Calibration of the thermometer was carried out from 34°C to 47°C with a water bath. There was a good linearity between the center wavelength of FBG and the tested temperature. The calibrated FBG thermometer was used to detect the temperature of the water bath with about 0.3°C interval. The standard error between the measured temperatures and the actual temperatures was 0.193°C .There was a little drift at different bending radius. The possible application of the temperature sensor was presented.
Keywords :
Bragg gratings; bending; biomedical equipment; calibration; fibre optic sensors; optical fibre cladding; patient diagnosis; thermometers; FBG sensor; bending radius; calibration; cladding mode suppression fiber; fiber Bragg grating; fiber sensing principle; medical applications; medical thermometry; optical fiber thermometer; temperature 34 degC to 47 degC; temperature sensor; Dielectrics; Fiber gratings; Fiber lasers; Gratings; Measurement by laser beam; calibration; characterization; fiber Bragg grating; temperature sensor;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electronics and Optoelectronics (ICEOE), 2011 International Conference on
Conference_Location :
Dalian
Print_ISBN :
978-1-61284-275-2
Type :
conf
DOI :
10.1109/ICEOE.2011.6013292
Filename :
6013292
Link To Document :
بازگشت