DocumentCode :
3212908
Title :
Design of a Calibration System for Measuring Probes
Author :
Hermann, Gyula
Author_Institution :
John von Neumann Fac. of Inf. Technol., Budapest Polytech., Budapest
fYear :
2008
fDate :
25-29 Feb. 2008
Firstpage :
59
Lastpage :
62
Abstract :
The probe is a major component of CMM. New high precision designs are available. In the corresponding literature little information can be found about the 3D calibration of these probes. The paper gives an overview of the novel probe designs, outlines the construction of a 3D calibration system, discusses it´s error sources, their contribution to the measuring uncertainty. A facility has been developed to calibrate analogue probes to an accuracy of 0,02 mum. A high precision three-axis translation stage moves the probe stylus and the position of the stage is determined by three mutually-orthogonal plane mirror laser inter-ferometer transducers. The coordinates indicated by the probe are compared with the co-ordinates indicated by the interferometers and the errors of the probe are then calculated. The working volume of is 100 times 100 times 100 mum. The full measurement strategy and design considerations are discussed and results are given to demonstrate its performance.
Keywords :
calibration; coordinate measuring machines; laser mirrors; light interferometers; probes; transducers; 3D calibration system; CMM; laser interferometer transducers; measuring probes; mutually-orthogonal plane mirror; Calibration; Capacitive sensors; Coordinate measuring machines; Displacement measurement; Information technology; Measurement uncertainty; Mirrors; Probes; Transducers; Tungsten;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Intelligent Engineering Systems, 2008. INES 2008. International Conference on
Conference_Location :
Miami, FL
Print_ISBN :
978-1-4244-2082-7
Electronic_ISBN :
978-1-4244-2083-4
Type :
conf
DOI :
10.1109/INES.2008.4481270
Filename :
4481270
Link To Document :