• DocumentCode
    3213323
  • Title

    Present Status of RF System for Medical Proton Synchrotron

  • Author

    Fang, Z. ; Endo, K. ; Egawa, K. ; Yamanaka, S. ; Hirashima, T. ; Fusato, T. ; Cho, Y.

  • Author_Institution
    KEK, Tsukuba, Japan, fang@post.kek.jp
  • fYear
    2005
  • fDate
    16-20 May 2005
  • Firstpage
    3185
  • Lastpage
    3187
  • Abstract
    The 200MeV proton synchrotron of circumference of 9.54m is being developed for medical radiotherapy. The rf system has been carried out with a wide bandwidth of frequency sweeping from 2.0MHz to 17.8MHz. The rf cavity is designed of a compact dimension and a high acceleration gradient. The high power test of the rf system has been successfully performed and maximal acceleration gradient of 60kV/m has been achieved. The experiments with rf feedback control system tracking to the exciting current of bending magnets are being studied In this paper, the recent progress of the rf system and test results will be presented in detail.
  • Keywords
    Acceleration; Bandwidth; Feedback control; Life estimation; Magnets; Performance evaluation; Protons; Radio frequency; Synchrotrons; System testing;