Title :
Field simulation of applicators for interstitial microwave hyperthermia
Author :
Schaller, G. ; Erb, J. ; Engelbrecht, Rainer ; Seegenschmiedt, H.M.
Author_Institution :
Lab. for High Frequency Technol., Erlangen-Nurnberg Univ., Germany
Abstract :
The electromagnetic field near clinically applied coaxial microwave applicators for interstitial hyperthermia is calculated using a finite integration algorithm program. Simulations of E-field and SAR distributions of different conventional applicators in a muscle phantom are presented and compared with measurements. The applied calculation method allows also design optimization and improvement of more sophisticated applicator types, e.g. those using a triaxial technique.<>
Keywords :
biomedical equipment; hyperthermia; microwave heating; muscle; radiation therapy; E-field; SAR distributions; applicators; clinically applied coaxial microwave applicators; design optimization; electromagnetic field; field simulation; finite integration algorithm program; interstitial microwave hyperthermia; muscle phantom; triaxial technique; Antenna measurements; Applicators; Catheters; Coaxial components; Electromagnetic heating; Frequency; Hyperthermia; Imaging phantoms; Impedance; Muscles;
Conference_Titel :
Microwave Symposium Digest, 1995., IEEE MTT-S International
Conference_Location :
Orlando, FL, USA
Print_ISBN :
0-7803-2581-8
DOI :
10.1109/MWSYM.1995.406045