    Validation of automatic CPAP leak algorithm

  • Author

    Grandjean-Thomsen, Nicolas ; Kang, Myeongsu ; Malouf, Gordon

  • Author_Institution
    ResMed Ltd., Sydney, NSW, Australia
  • fYear
    2013
  • fDate
    3-7 July 2013
  • Firstpage
    6873
  • Lastpage
    6876
  • Abstract
    Obstructive Sleep Apnoea (OSA) is a prevalent disease. Reliable estimation of respiratory parameters by devices used to treat OSA is important for therapy initiation and maintenance. This is achieved by estimating patient flow in the presence of inadvertent leak from the total flow measured by the device. A method of validating the patient flow estimation of an Automatic Continuous Positive Airway Pressure (APAP) device is described. Novel techniques of a multi-composite simulant head and recorded patient flows are used. The APAP device tested was shown to reliably estimate patient flow across a range of therapy pressures, leak conditions and breath types.
  • Keywords
    biomedical equipment; diseases; patient treatment; pneumodynamics; sleep; APAP device; Automatic Continuous Positive Airway Pressure device; Obstructive Sleep Apnoea; automatic CPAP leak algorithm; breath type; leak condition; multicomposite simulant head; patient flow estimation; prevalent disease; respiratory parameter; therapy initiation; therapy maintenance; therapy pressure; total flow; Fluid flow measurement; Head; Magnetic heads; Medical treatment; Morphology; Pistons; Sleep apnea;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
  • Conference_Location
    Osaka
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2013.6611137
  • Filename
    6611137