• DocumentCode
    3219222
  • Title

    Designing a flowmeter for hydrocephalus patients

  • Author

    Lustig, N. ; Liu, S. ; Thomas, G.A.

  • Author_Institution
    Dept. of Phys., New Jersey Inst. of Technol., Newark, NJ, USA
  • fYear
    2005
  • fDate
    2-3 April 2005
  • Firstpage
    252
  • Lastpage
    253
  • Abstract
    We built a small flowmeter system to measure low flow rates. The flowmeter indicated different flow rates by producing a voltage measure. The flowmeter is sensitive at low flow rates (below 1.8 ml/min) but there is a saturation point at higher flow rates (above 2 ml/min). This flowmeter could be used to develop a smart shunt for patients with hydrocephalus.
  • Keywords
    biological fluid dynamics; biomedical measurement; diseases; flowmeters; neurophysiology; cerebrospinal fluid; flow rate measurement; flowmeter design; hydrocephalus patient; neurological condition; saturation point; shunt system; voltage measure; Cranial pressure; Fluid flow measurement; Implants; Physics; Plastics; Power supplies; Springs; Valves; Voltage measurement; Voltmeters;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference, 2005. Proceedings of the IEEE 31st Annual Northeast
  • Print_ISBN
    0-7803-9105-5
  • Electronic_ISBN
    0-7803-9106-3
  • Type

    conf

  • DOI
    10.1109/NEBC.2005.1432015
  • Filename
    1432015