• DocumentCode
    3226088
  • Title

    Nonlinear parameter imaging to characterize HIFU ablation: Preliminary in vitro results in porcine liver

  • Author

    Varray, François ; Chenot, Jérémy ; Basset, Olivier ; Tortoli, Piero ; Melodelima, David ; Cachard, Christian

  • Author_Institution
    CREATIS, Univ. de Lyon, Villeurbanne, France
  • fYear
    2011
  • fDate
    18-21 Oct. 2011
  • Firstpage
    1361
  • Lastpage
    1363
  • Abstract
    In nonlinear US propagation, the B/A parameter allow to quantify the nonlinear distortion of the US pulse and the apparition of the different harmonics. Previous studies highlighted the interest of the nonlinear parameter in tissue characterization or parametric imaging. In this paper, an extended comparative method (ECM) is used to image the nonlinear parameter of in vitro porcine liver. The nonlinear parameter is imaged before and after a high intensity focused ultrasound (HIFU) treatment. The resulting maps highlight the region of the lesion and show the potential of B/A imaging in HIFU applications.
  • Keywords
    biological tissues; biomedical ultrasonics; liver; ultrasonic propagation; B-A imaging; B-A parameter; HIFU ablation; HIFU applications; HIFU treatment; extended comparative method; high intensity focused ultrasound treatment; in vitro porcine liver; lesion region; nonlinear distortion; nonlinear parameter imaging; nonlinear ultrasound propagation; preliminary in vitro results; tissue characterization; ultrasound pulse; Electronic countermeasures; Frequency conversion; Imaging; Lesions; Liver; Medical treatment; Ultrasonic imaging; HIFU; Nonlinear parameter; tissue characterization;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2011 IEEE International
  • Conference_Location
    Orlando, FL
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-1253-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2011.0336
  • Filename
    6293206