  • Title

    Design and development of a low cost EMG signal acquisition system using surface EMG electrode

  • Author

    Poo, T.S. ; Sundaraj, K.

  • Author_Institution
    Sch. of Mechatron. Eng., Univ. Malaysia Perlis (UniMAP), Arau, Malaysia
  • fYear
    2010
  • fDate
    6-9 Dec. 2010
  • Firstpage
    24
  • Lastpage
    27
  • Abstract
    Electromyogram or EMG signal is a very small signal; it requires a system to enhance for display purpose or for further analysis process. This paper presents the development of low cost physiotherapy EMG signal acquisition system with two channel input. In the acquisition system, both input signals are amplified with a differential amplifier and undergo signal pre-processing to obtain the linear envelope of EMG signal. Obtained EMG signal is then digitalized and sent to the computer to be plotted.
  • Keywords
    biomedical electrodes; biomedical electronics; differential amplifiers; electromyography; medical signal detection; medical signal processing; patient treatment; signal processing equipment; EMG signal acquisition system; differential amplifier; electromyogram; low-cost physiotherapy system; signal preprocessing; surface EMG electrode; two-channel input; Computers; Electrodes; Electromyography; Instruments; Low pass filters; Muscles; Nerve fibers; Acquisition System; Differential Amplifier; Electromyogram; Linear Envelope; Surface EMG Electrode;