DocumentCode :
3229335
Title :
Ultrasound-based Speckle Tracking for 3D Strain estimation of the Arterial wall — An experimental validation study in a tissue mimicking phantom
Author :
Larsson, Mats ; Kremer, Florence ; Heyde, Brecht ; Brodin, Lars-Ake ; D´hooge, J.
Author_Institution :
Dept. of Cardiovascular Diseases, Catholic Univ. of Leuven, Leuven, Belgium
fYear :
2011
fDate :
18-21 Oct. 2011
Firstpage :
725
Lastpage :
728
Abstract :
Arterial stiffness is an important risk factor for cardiovascular disease. As such, ultrasound-based methods have been proposed to assess arterial strain as a measure of stiffness. The aim of the current study was to validate our recently proposed speckle tracking (ST) algorithm to estimate the in-plane wall strain tensor in an experimental setup. Three polyvinyl alcohol phantoms mimicking the carotid artery were constructed with different mechanical properties (2, 3 and 4 freeze-thaw cycles). The phantoms were connected to a pump, programmed to generate carotid flow profiles at peak flows of 7, 14, 21, 28 and 35 ml/s. Long and short-axis ultrasound images were obtained using a Vivid7 Dimension system. Radial, longitudinal and circumferential strains were estimated using the ST algorithm (kernel size: 2.7λ×2λ, normalized cross-correlation; spline inter-polation for subsample motion estimation; 40% window overlap). Sonomicrometry was used to acquire reference values of strain in the phantoms. Good agreement was found between the estimated radial, longitudinal and circumferential strain and the acquired reference strain. The correlation between estimated mean peak strain values and reference peak strain values was r = 0.92 (p <;0.001) for radial strain, r = 0.72 (p = 0.006) for longitudinal strain and r = 0.91 (p<;0.001) for circumferential strain.
Keywords :
biomechanics; biomedical measurement; biomedical ultrasonics; blood vessels; cardiovascular system; diseases; elastic constants; image sampling; interpolation; medical image processing; motion estimation; phantoms; speckle; strain measurement; ultrasonic imaging; 3D strain estimation; Vivid7 Dimension system; arterial stiffness measurement; arterial wall; cardiovascular disease; carotid flow profiles; circumferential strains; freeze-thaw cycles; in-plane wall strain tensor; long-axis ultrasound images; longitudinal strains; mechanical properties; polyvinyl alcohol phantoms; pump; radial strains; short-axis ultrasound images; sonomicrometry; spline interpolation; subsample motion estimation; tissue mimicking phantom; ultrasound-based speckle tracking; Carotid arteries; Crystals; Phantoms; Speckle; Strain; Strain measurement; Ultrasonic imaging; Arterial strain; Carotid artery; Speckle tracking; Vessel phantom;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location :
Orlando, FL
ISSN :
1948-5719
Print_ISBN :
978-1-4577-1253-1
Type :
conf
DOI :
10.1109/ULTSYM.2011.0176
Filename :
6293377
Link To Document :
بازگشت