• DocumentCode
    3230699
  • Title

    Development of a liquid oxygen density reference system

  • Author

    Li, Yayu

  • Author_Institution
    Beijing Res. Inst. of Test Technol., China Nat. Space Admin., China
  • fYear
    1997
  • fDate
    29 Sep-2 Oct 1997
  • Firstpage
    153
  • Lastpage
    162
  • Abstract
    The development of a liquid oxygen density reference system is used in the measurement of liquid oxygen density. The system consists mainly of liquid oxygen container, special purpose thermostat, pressure control device, buoy, evacuation system, electronic balance measurement system, tracking and commanding system of chip microprocessor, and PC. The analysis of measurement uncertainty for the liquid oxygen density reference system indicates that the system error is ±0.02%. In the light of random error possibly resulting from measurements, the measurement uncertainty of liquid oxygen density for the system is ±0.05%. The measurement of liquid oxygen density will be more correct, reliable and convenient for the liquid oxygen density reference system, when compared with sensor of traditional liquid oxygen density measurement
  • Keywords
    balances; cryogenics; density measurement; measurement errors; oxygen; cryogenics; densimeters; electronic balance measurement system; evacuation system; liquid oxygen density reference system; measurement uncertainty; pressure control device; random error; system error; thermostat; tracking; Containers; Density measurement; Error correction; Measurement uncertainty; Microprocessors; Pressure control; Pressure measurement; Semiconductor device measurement; Sensor systems; Thermostats;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation in Aerospace Simulation Facilities, 1997. ICIASF '97 Record., International Congress on
  • Conference_Location
    Pacific Grove, CA
  • Print_ISBN
    0-7803-4167-8
  • Type

    conf

  • DOI
    10.1109/ICIASF.1997.644677
  • Filename
    644677