• DocumentCode
    323193
  • Title

    Progress in the development of large-area modular 64×64 CdZnTe imaging arrays for nuclear medicine

  • Author

    Matherson, K.J. ; Barber, H.B. ; Barrett, H.H. ; Eskin, J.D. ; Dereniak, E.L. ; Marks, D.G. ; Woolfenden, J.M. ; Young, E.T. ; Augustine, F.L.

  • Author_Institution
    Opt. Sci. Center, Arizona Univ., Tucson, AZ, USA
  • fYear
    1997
  • fDate
    9-15 Nov 1997
  • Firstpage
    276
  • Abstract
    Previous efforts by our group have demonstrated the potential of hybrid semiconductor detector arrays for use in gamma-ray imaging applications. In this paper, we describe progress in the development of a prototype imaging system consisting of a 64×64-pixel CdZnTe detector array mated to a multiplexer readout circuit that was custom designed for our nuclear medicine application. The detector array consists of a 0.15 cm thick slab of CdZnTe which has a 64×64 array of 380 μm square pixel electrodes on one side produced by photolithography; the other side has a continuous electrode biased at -150 V. Electrical connections between the detector electrodes and corresponding multiplexer bump pads are made with indium bump bonds. Although the CdZnTe detector arrays characterized in this paper are room-temperature devices, a slight amount of cooling is necessary to reduce thermally generated dark current in the detectors. Initial tests show that this prototype imager functions well with more than 90% of its pixels operating. The device is an excellent imager; phantom images have a spatial resolution of 1.5 mm, limited by the collimator bore
  • Keywords
    II-VI semiconductors; brain; cadmium compounds; gamma-ray detection; semiconductor counters; single photon emission computed tomography; zinc compounds; -150 V; CdZnTe; SPECT; brain imaging; collimator bore; continuous electrode; gamma-ray imaging; hybrid semiconductor detector arrays; large-area modular imaging arrays; multiplexer bump pads; multiplexer readout circuit; nuclear medicine; phantom images; prototype imaging system; single-photon emission computed tomography; square pixel electrodes; thermally generated dark current; Biomedical electrodes; Circuits; Gamma ray detection; Gamma ray detectors; Multiplexing; Nuclear imaging; Nuclear medicine; Optical imaging; Prototypes; Sensor arrays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1997. IEEE
  • Conference_Location
    Albuquerque, NM
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-4258-5
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1997.672584
  • Filename
    672584