• DocumentCode
    3232177
  • Title

    Evaluation of 3D point spread function of a semi-ellipsoidal ultrasound computer tomography system

  • Author

    Ruiter, N.V. ; Dapp, R. ; Zapf, M. ; Gemmeke, H.

  • Author_Institution
    Inst. for Data Process. & Electron., Karlsruhe Inst. of Technol., Karlsruhe, Germany
  • fYear
    2011
  • fDate
    18-21 Oct. 2011
  • Firstpage
    242
  • Lastpage
    245
  • Abstract
    The design of 3DUSCT II was optimized with the aim of nearly isotropic and approx. spatially invariant 3D point spread function (PSF). To evaluate experimentally the 3DPSF, a thin wire was twisted into an irregular helix and imaged with the USCT system. The PSF was measured at 55 points along the wire as 2D profiles of the full width at half maximum (FWHM) in a plane perpendicular to the local direction of the thread. The mean FWHM over all profiles was 0.66 mm. The standard deviation (STD) of the FWHM profiles as measure of the isotropy had a mean of 0.18 mm. The variability of the mean FWHM over the points shows no systematic trend over the imaged space.
  • Keywords
    acoustic tomography; biomedical ultrasonics; cancer; computerised tomography; medical image processing; transfer functions; 3D point spread function; 3DUSCT II design optimisation; USCT imaging; approximately spatially invariant 3D PSF; irregular helical wire; nearly isotropic 3D PSF; semiellipsoidal ultrasound computer tomography system; Apertures; Breast; Image reconstruction; Imaging; Transducers; Ultrasonic imaging; Wires;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2011 IEEE International
  • Conference_Location
    Orlando, FL
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-1253-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2011.0059
  • Filename
    6293524