• DocumentCode
    3235058
  • Title

    Development of a high-definition intravascular ultrasound imaging system and catheter

  • Author

    Waters, Kendall R. ; Bautista, Rick ; Zelenka, Robert ; Masters, Donald ; Reynolds, J. Steve ; Nelson, Steve ; Lam, Duc H. ; Neville, Ron ; Moore, Thomas C.

  • Author_Institution
    Silicon Valley Med. Instrum., Inc., Fremont, CA, USA
  • fYear
    2011
  • fDate
    18-21 Oct. 2011
  • Firstpage
    1762
  • Lastpage
    1765
  • Abstract
    We have developed a high-frequency intravascular ultrasound (IVUS) system and catheter that is capable of imaging through a blood field at 40 MHz and 60 MHz. Representative transducers exhibit >;50 MHz center frequency, >;50 % fractional bandwidth, and a -20 dB pulse duration of <;80 ns. A series of bench, ex vivo, non-clinical, and clinical studies demonstrate that benefits of the IVUS imaging system and catheter include improved spatial and contrast resolution. The 60 MHz IVUS images appear to enable improved detection of lipid-rich plaque and thrombus, which can potentially impact treatment decisions.
  • Keywords
    biomedical transducers; biomedical ultrasonics; cardiology; catheters; diseases; patient treatment; IVUS imaging system; blood field; catheter; fractional bandwidth; frequency 40 MHz; frequency 60 MHz; high-definition intravascular ultrasound imaging system; high-frequency intravascular ultrasound system; lipid-rich plaque; potentially impact treatment decisions; pulse duration; spatial contrast resolution; thrombus; transducers; Arteries; Catheters; Heart; Imaging; Time domain analysis; Transducers; Wires; clinical; coronary heart disease; intravascular ultrasound;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2011 IEEE International
  • Conference_Location
    Orlando, FL
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-1253-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2011.0440
  • Filename
    6293668