• DocumentCode
    3235446
  • Title

    Multiple-frequency calibration of two-dimensional hydrophone arrays for instantaneous beam width measurements of shock-wave lithotripters

  • Author

    Gross, Daniel ; Filoux, Erwan ; Ketterling, Jeffrey A.

  • Author_Institution
    Riverside Res., Lizzi Center for Biomed. Eng., New York, NY, USA
  • fYear
    2011
  • fDate
    18-21 Oct. 2011
  • Firstpage
    1660
  • Lastpage
    1663
  • Abstract
    Characterizations of electrohydraulic lithotripters (EHLs) are usually limited to single-point measurements because hydrophone arrays are not readily available, or are prohibitively expensive. Furthermore, beamwidth measurements of peak pressure values made with single-point hydrophones are inaccurate. This inaccuracy comes from the high variability of the acoustic field and the fact that several shocks must be averaged to create a single beamwidth profile when using single-element hydrophones. Previous work demonstrated that characterizing the beam of an EHL with a single-point measurement technique led to an overestimate error of 25% compared to an instantaneous beamwidth measurement using a linear hydrophone array. In this paper, the improvement of our disposable, repeatable, hydrophone array is presented. The geometry was changed to that of a two-dimensional hydrophone array, which allows for the measurement of the instantaneous acoustic field of an EHL and its variation in 2-D. Modifications were made to the hydrophone fabrication methods which improved sensitivity and performance. Finally, a multiple-frequency calibration technique was implemented, allowing absolute pressures to be obtained for each hydrophone element.
  • Keywords
    acoustic variables measurement; biomedical transducers; biomedical ultrasonics; calibration; hydrophones; ultrasonic therapy; ultrasonic transducer arrays; 2D hydrophone arrays; EHL; acoustic field variability; beamwidth profile; disposable hydrophone array; electrohydraulic lithotripters; hydrophone fabrication methods; instantaneous beam width measurements; instantaneous beamwidth measurement; linear hydrophone array; multiple frequency calibration; peak pressure beamwidth measurements; repeatable hydrophone array; shock wave lithotripters; single element hydrophones; Acoustic measurements; Acoustics; Arrays; Biomedical measurements; Calibration; Sensitivity; Sonar equipment;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2011 IEEE International
  • Conference_Location
    Orlando, FL
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-1253-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2011.0413
  • Filename
    6293688