• DocumentCode
    3235819
  • Title

    New design of double-core QCM system

  • Author

    Jiang, Haifeng ; Bo, Yuming

  • Author_Institution
    Inst. of Autom., NanJing Univ. of Sci. & Technol., Nanjing, China
  • fYear
    2010
  • fDate
    8-11 May 2010
  • Firstpage
    921
  • Lastpage
    924
  • Abstract
    QCM(Quartz Crystal Microbalance) is a piezoelectricity sensor with nanogram precision. However twice measurements of no-load and a load are necessary in the tradition, which will result in frequency errors when environments(such as temperature and humidity) changes. This paper affords a new design of double-core QCM for solving the time-delay problem in the QCM traditional measuring method. Reference QCM is introduced, QCM signal drive and acquisition circuit based on correlation theory is designed and frequency characteristic curve of double-core QCM is analyzed.
  • Keywords
    delays; microbalances; piezoelectric transducers; QCM signal drive; acquisition circuit; correlation theory; double-core QCM system; frequency characteristic curve; frequency errors; piezoelectricity sensor; quartz crystal microbalance; time-delay problem; Circuits; Communication system control; Frequency measurement; Humidity measurement; Materials testing; Resonant frequency; Signal design; Temperature control; Time measurement; Vibration measurement; EEACC:7320G; PACC:8280T; QCM; correlation theory; double-core; frequency characteristic curve;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave and Millimeter Wave Technology (ICMMT), 2010 International Conference on
  • Conference_Location
    Chengdu
  • Print_ISBN
    978-1-4244-5705-2
  • Type

    conf

  • DOI
    10.1109/ICMMT.2010.5525151
  • Filename
    5525151