DocumentCode
3244148
Title
Optimal signal recovery from interleaved FMRI data
Author
Parker, David ; Gerraty, Raphael T. ; Razlighi, Qolamreza R.
Author_Institution
Biomed. Eng., Columbia Univ., Columbia, NY, USA
fYear
2015
fDate
16-19 April 2015
Firstpage
1372
Lastpage
1375
Abstract
Due to the nature of fMRI acquisition protocols, slices in the plane of acquisition are not acquired simultaneously or sequentially, and therefore are temporally misaligned with each other. Slice timing correction (STC) is a critical preprocessing step that corrects for this misalignment. STC is applied in all major software packages. To date, little effort has gone towards assessing the optimal method of STC. In this study, we examine the most popular methods of STC, and propose a new optimal method based on the fundamental properties of sampling theory. We evaluate this method using 20 simulated fMRI data and demonstrate the utility of STC in general as well as the superiority of the proposed method in comparison to existing ones.
Keywords
biomedical MRI; medical signal processing; STC optimal method; fMRI acquisition protocol; interleaved FMRI data; optimal signal recovery; sampling theory; slice timing correction; software package; Delays; Frequency-domain analysis; Interpolation; Kernel; Noise; Shape; EPI; Interleaved acquisition; Interpolation; Slice timing correction; fMRI;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Imaging (ISBI), 2015 IEEE 12th International Symposium on
Conference_Location
New York, NY
Type
conf
DOI
10.1109/ISBI.2015.7164131
Filename
7164131
Link To Document