• DocumentCode
    3275060
  • Title

    Design of Gyrotron FU CW VI for 600 MHz DNP-NMR experiment

  • Author

    Glyavin, M. ; Idehara, T. ; Khizhnyak, V. ; Luchinin, A. ; Manuilov, V. ; LaAgusu ; Ogawa, I. ; Saito, T. ; Takahashi, H. ; Fujiwara, T.

  • Author_Institution
    Inst. of Appl. Phys., Russian Acad. of Sci., Nizhny Novgorod
  • fYear
    2008
  • fDate
    15-19 Sept. 2008
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    For the application of gyrotron to accurate measurement of DNP-NMR, frequency continuous tunability is desired in 2 GHz range at least. The demountable Gyrotron FU CW VI tube based on 15 T helium free superconducting magnet are under development. The gyrotron operation is in CW mode at the fundamental cyclotron harmonic. The output power is 100 W in the whole frequency tunable range and the center frequency is 394.6 GHz. The gyrotron will be applied to 600 MHz DNP-NMR experiments.
  • Keywords
    gyrotrons; DNP-NMR experiment; Gyrotron FU CW VI; frequency continuous tunability; fundamental cyclotron harmonic; helium free superconducting magnet; magnetic flux density 15 T; Coaxial components; Conductors; Eigenvalues and eigenfunctions; Frequency; Gyrotrons; Impedance; Magnetic fields; Nuclear magnetic resonance; Physics; Tuning;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Infrared, Millimeter and Terahertz Waves, 2008. IRMMW-THz 2008. 33rd International Conference on
  • Conference_Location
    Pasadena, CA
  • Print_ISBN
    978-1-4244-2119-0
  • Electronic_ISBN
    978-1-4244-2120-6
  • Type

    conf

  • DOI
    10.1109/ICIMW.2008.4665547
  • Filename