DocumentCode :
3286796
Title :
Development of Online Measuring Instrument for Water-in-Oil percentage Based on MCU
Author :
Nan, Xie ; Haibo, Zhang ; Shenyong, Gao ; Yan, Ma ; Weimin, Chen
Author_Institution :
Dept. of Comput. & Inf. Eng., Zhejiang Water Conservancy & Hydropower Coll., Hangzhou, China
Volume :
3
fYear :
2009
fDate :
15-17 May 2009
Firstpage :
563
Lastpage :
565
Abstract :
Water may access to hydraulic and lubricant system due to various reasons, so the lubricant oil in hydraulic system will be out of work for high water-in-oil percentage. Traditional measuring instruments for water-in-oil percentage have slower response-time and higher costs. The principle of water-in-oil percentage detection, capacitance to voltage signal conversion and the structure of measurement system are proposed in this paper. It is demonstrated that the detection device which uses the mixed technology of real-time MCU, XE2004 conversion chip and smart sensor can quickly and precisely measure water-in-oil percentage and avoid the potential hazards.
Keywords :
capacitive sensors; hydraulic systems; intelligent sensors; lubricating oils; MCU; XE2004 conversion chip; capacitance-voltage signal conversion; hydraulic system; lubricant oil; measurement system structure; online measuring instrument; smart sensor; water-in-oil percentage detection; Capacitance measurement; Costs; Hazards; Hydraulic systems; Instruments; Intelligent sensors; Lubricants; Petroleum; Semiconductor device measurement; Voltage; MCU; XE2004; capacitance to voltage convention; online measurement; water-in-oil percentage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Information Technology and Applications, 2009. IFITA '09. International Forum on
Conference_Location :
Chengdu
Print_ISBN :
978-0-7695-3600-2
Type :
conf
DOI :
10.1109/IFITA.2009.365
Filename :
5232187
Link To Document :
بازگشت