• DocumentCode
    3287724
  • Title

    Sectional Ultrasonic Irradiator for Intracavitary Treatment

  • Author

    Belavskaya, S.V. ; Kuchma, I.A. ; Lisitsyna, L.I. ; Adonev, V.G. ; Firsova, K.F.

  • Author_Institution
    Novosibirsk State Techn. Univ., Novosibirsk
  • fYear
    2006
  • fDate
    26-28 Sept. 2006
  • Firstpage
    32
  • Lastpage
    37
  • Abstract
    In this paper experimental results on sectional ultrasonic irradiator (USI), which is implemented on several piezoplates located parallel to each other on lateral surface of cylindrical body, are presented. It is shown that with section number increase from 1 up to 4, uniformity of irradiation intensity across USI sharply increases. The further increase in number of sections increases the uniformity more poorly. It is noted, that six-section USI provides fluctuation of intensity across radiator only on 10-15 %.
  • Keywords
    biomedical equipment; biomedical ultrasonics; fluctuations; patient treatment; piezoelectric devices; ultrasonic devices; cylindrical body; intracavitary treatment; irradiation intensity uniformity; piezoplates; sectional ultrasonic irradiator; ultrasonic intensity fluctuation; Biological tissues; Catheters; Cows; Design methodology; Fluctuations; Research and development; Surface treatment; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electronic Instrument Engineering, 2006. APEIE '06. 8th International Conference on Actual Problems of
  • Conference_Location
    Novosibirsk
  • Print_ISBN
    5-7782-0662-3
  • Type

    conf

  • DOI
    10.1109/APEIE.2006.4292419
  • Filename
    4292419