• DocumentCode
    3294344
  • Title

    Phased array beam forming from a circular array: applications to imaging of coronary arteries

  • Author

    Donnell, M.O.

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Sci., Michigan Univ., Ann Arbor, MI, USA
  • fYear
    1991
  • fDate
    8-11 Dec 1991
  • Firstpage
    637
  • Abstract
    A catheter based, circular phased array system is presented for side-looking, high-resolution imaging of coronary arteries. The system is capable of full dynamic focus over a large depth of field using only a single connection between the catheter tip and the imaging console. The results of detailed simulations are presented, demonstrating the capabilities of the method
  • Keywords
    acoustic imaging; biomedical ultrasonics; cardiology; patient diagnosis; ultrasonic transducers; catheter tip; circular phased array system; coronary arteries; high-resolution imaging; imaging console; medical diagnostic imaging; Application software; Arteries; Catheters; Focusing; Frequency; High-resolution imaging; Lenses; Phased arrays; Ultrasonic imaging; Wires;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1991. Proceedings., IEEE 1991
  • Conference_Location
    Orlando, FL
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1991.234240
  • Filename
    234240