DocumentCode :
3298271
Title :
Monolithic neuro-Stimulus chip for retinal implants
Author :
Li, Kai ; Zhang, Xu ; Pei, WeiHua ; Huang, Beiju ; Wang, ShuJing ; Wu, Huijuan ; Wang, Kai ; Li, Xiaoxin ; Chen, HongDa
Author_Institution :
Dept. of Epidemiology & Biostat., Peking Univ., Beijing
fYear :
2009
fDate :
9-11 April 2009
Firstpage :
1
Lastpage :
4
Abstract :
A prototype neuro-stimulus chip for retinal implants in blind patients affected by Age-related Macular Degeneration (AMD) or Retinitis Pigmentosa (RP) is presented in this paper. This retinal prosthetic chip was designed to replace the degenerated photoreceptor cells, and in order to stimulate directly the remaining healthy layers of retinal neurons. The current stimulus circuits are monolithic integrated with photodiodes (PD) array, which can convert the illumination on the eyes into 3-bit bi-phasic electrical pulses. In addition, a novel charge cancellation circuit is used to discharge the electrodes for medical safty. The prototype chip is designed and fabricated in HJTC 0.18 mum N-well CMOS 1P6M Mix-signal process, with a plusmn 2.5 V dual voltage power supply.
Keywords :
CMOS digital integrated circuits; eye; photodiodes; prosthetics; Age-related Macular Degeneration; Retinitis Pigmentosa; monolithic neurostimulus chip; photodiodes array; photoreceptor cells; retinal implants; retinal neurons; Circuits; Implants; Neurons; Partial discharges; Photodiodes; Photoreceptors; Pigmentation; Prosthetics; Prototypes; Retina;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Complex Medical Engineering, 2009. CME. ICME International Conference on
Conference_Location :
Tempe, AZ
Print_ISBN :
978-1-4244-3315-5
Electronic_ISBN :
978-1-4244-3316-2
Type :
conf
DOI :
10.1109/ICCME.2009.4906638
Filename :
4906638
Link To Document :
بازگشت