• DocumentCode
    3298886
  • Title

    SiliPET: design of an ultra high resolution small animal PET scanner based on stacks of semiconductor detectors

  • Author

    Auricchio, Natalia ; Cesca, Nicola ; Domenico, Giovanni Di ; Moretti, Elena ; Sabba, Nicola ; Gambaccini, Mauro ; Zavattini, Guido ; Andritschke, Robert ; Kanbach, Gottfried ; Schopper, Florian

  • Author_Institution
    Dept. of Phys., Ferrara Univ.
  • Volume
    5
  • fYear
    2005
  • fDate
    23-29 Oct. 2005
  • Firstpage
    3010
  • Lastpage
    3013
  • Abstract
    We have studied with Monte Carlo simulations, using the EGSnrc code, a new scanner for small-animal positron emission tomography (PET) based on stacks of double sided semiconductor detectors. In small animal PET imaging (mice) Compton scattering within the animal itself is not an issue and therefore registration of the energy information is not necessary. Low Z materials can therefore be studied for designing a high performance scanner. Each stack is composed of planar detectors with dimension 60times60times1 mm3 and 128 orthogonal strips on both sides to read the two coordinates of the interaction, the third coordinate being the detector number in the stack. Coincidence events are recorded only if two plane detectors in two different stacks registered an energy deposition over threshold. In this way we achieve a precise determination of the interaction point of the two 511 keV photons, overcoming three of the crucial scanner limits: depth of interaction measurement thus eliminating parallax error; inter-detector multiple interactions by requiring a single hit; reduced distance between the detectors reduces significantly effects due to the non colinearity of the emitted gamma rays. The reduced dimensions of the scanner also improve the sold angle coverage resulting in a high sensitivity. This suggests that this is a promising new approach for small animal PET imaging. The sinogram profile for a pointlike source of 18F in a 3 cm diameter water phantom is 0.52 mm FWHM. Due to the large solid angle coverage and relatively high Compton efficiency, the simulated sensitivity at the center of the FOV is 5.1% with 4 stacks each 4 cm thick placed in a box-like configuration. Preliminary results of a proof of principle measurement done with the MEGA advanced Compton imager using a sime1 mm diameter 22Na source showed a focal plane ray tracing FWHM of 1 mm
  • Keywords
    Monte Carlo methods; coincidence techniques; position sensitive particle detectors; positron emission tomography; silicon radiation detectors; 1 mm; 3 cm; 60 mm; 18F; Compton efficiency; EGSnrc code; MEGA advanced Compton imager; Monte Carlo simulations; SiliPET; coincidence events; double sided silicon strip detectors; gamma rays; interdetector multiple interactions; parallax error; planar detectors; pointlike source; sinogram profile; small animal PET imaging Compton scattering; small animal PET scanner; small-animal positron emission tomography; water phantom; Animals; Detectors; Electromagnetic scattering; Event detection; High-resolution imaging; Mice; Particle scattering; Positron emission tomography; Semiconductor materials; Strips;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2005 IEEE
  • Conference_Location
    Fajardo
  • ISSN
    1095-7863
  • Print_ISBN
    0-7803-9221-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2005.1596964
  • Filename
    1596964