DocumentCode
3301257
Title
A Low Cost Sensing Device to Detect Cardiac Timing and Function
Author
Anh Dinh ; Seokbum Ko ; Neary, P. ; Quarrie, D.M.
Author_Institution
Dept. of Electr. & Comput. Eng., Univ. of Saskatchewan, Saskatoon, SK, Canada
fYear
2012
fDate
14-16 May 2012
Firstpage
101
Lastpage
104
Abstract
This paper presents a design of a device to detect cardiac timing required for diagnosis of cardiac deficiencies. This low cost device can have an accuracy which is comparable to the expensive echocardiography when combined with the ECG signal. The device includes an ECG sensor, a microphone to record phonocardiogram, and a tri-axis accelerometer to detect the seismic motion of the heart. Data are collected in a logger to view in real time or store in a file for later analysis. Experiment results have shown a working prototype to provide accurate waveforms of the seism cardiogram to identify critical points in a cardiac cycle to millisecond accuracy.
Keywords
accelerometers; biomedical equipment; electrocardiography; medical signal detection; microphones; ECG sensor; cardiac deficiency diagnosis; cardiac timing detection; cardiogram; low cost device; microphone; phonocardiogram; seismic motion; triaxis accelerometer; Accelerometers; Echocardiography; Electrocardiography; Heart; Sensors; Timing; cardiac cycle; seismocardiogram; sensors; tri-axis accelerometer;
fLanguage
English
Publisher
ieee
Conference_Titel
Mixed-Signals, Sensors and Systems Test Workshop (IMS3TW), 2012 18th International
Conference_Location
Taipei
Print_ISBN
978-1-4673-1925-6
Type
conf
DOI
10.1109/IMS3TW.2012.29
Filename
6298725