DocumentCode :
3302325
Title :
Merging visible and invisible: two Camera-Augmented Mobile C-arm (CAMC) applications
Author :
Navab, N. ; Bani-Kashemi, A. ; Mitschke, M.
Author_Institution :
Siemens Corp. Res. Inc., Princeton, NJ, USA
fYear :
1999
fDate :
1999
Firstpage :
134
Lastpage :
141
Abstract :
This paper presents the basic concept of CAMC and some of its applications. A CCD camera is attached to a mobile C-arm fluoroscopy X-ray system. Both optical and X-ray imaging systems are calibrated in the same coordinate system in an off-line process. The new system is able to provide X-ray and optical images simultaneously. The CAMC framework has great potential for medical augmented reality. We briefly introduce two new CAMC applications to the augmented reality research community. The first application aims at merging video images with a pre-computed tomographic reconstruction of the 3D volume of interest. This is a logical continuation of our work on 3D reconstruction using a CAMC (1999). The second approach is a totally new CAMC design where using a double mirror system and an appropriate calibration procedure the X-ray and optical images are merged in real-time. This new system enables the user to see an optical image, an X-ray image, or an augmented image where both visible and invisible are combined in real-time. The paper is organized in two independent sections describing each of the above. Experimental results are provided at the same time as the methods and apparatus are described for each section
Keywords :
CCD image sensors; X-ray imaging; augmented reality; computerised tomography; image reconstruction; medical image processing; 3D volume of interest; CAMC; CCD camera; Camera-Augmented Mobile C-arm; X-ray imaging; augmented reality; calibration procedure; coordinate system; double mirror system; medical augmented reality; mobile C-arm fluoroscopy X-ray system; optical imaging; real-time systems; tomographic reconstruction; video images; Augmented reality; Biomedical imaging; Biomedical optical imaging; Charge coupled devices; Charge-coupled image sensors; Image reconstruction; Merging; Real time systems; Tomography; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Augmented Reality, 1999. (IWAR '99) Proceedings. 2nd IEEE and ACM International Workshop on
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7695-0359-4
Type :
conf
DOI :
10.1109/IWAR.1999.803814