DocumentCode :
3303977
Title :
Aortic pressure with asynchronous ventricular assist
Author :
Gu, M. ; Gaumond, R. ; Snyder, A. ; Gardner, J.
Author_Institution :
Dept. of Bioeng., Pennsylvania State Univ., University Park, PA, USA
Volume :
2
fYear :
1999
fDate :
36434
Abstract :
The authors have examined the relation between Pao (aortic pressure) and parameters of an asynchronous VAD (ventricular assist device) in a lumped-parameter simulation. While left ventricular volume was greater than a threshold level Vth, the 50 ml VAD chamber filled passively at 100 ml/sec. The VAD emptied at a rate of 400 ml/sec when filled. The ventricular elastance (1/C) followed a half-rectified sine waveform. Over a wide range of heart rates determined by Vth, VAD pumping was entrained synchronous to ventricular pumping by this protocol. With the VAD included, Pao(t) waveform complexity increased, with generally more high-frequency signal energy. This has implications for baroreceptor reflex function in VAD patients
Keywords :
artificial organs; cardiology; haemodynamics; physiological models; aortic pressure; asynchronous ventricular assist; baroreceptor reflex function; cardiovascular model; half-rectified sine waveform; high-frequency signal energy; left ventricular volume; lumped-parameter simulation; passive chamber filling; ventricular elastance; Baroreflex; Heart rate; Protocols;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
[Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
Conference_Location :
Atlanta, GA
ISSN :
1094-687X
Print_ISBN :
0-7803-5674-8
Type :
conf
DOI :
10.1109/IEMBS.1999.803913
Filename :
803913
Link To Document :
بازگشت