• DocumentCode
    3309587
  • Title

    A focal plane array system for clinical infra red imaging

  • Author

    Ring, E.F.J. ; Minchinton, M.A.B. ; Elvins, D.M.

  • Author_Institution
    R. Nat. Hosp. for Rheumatic Diseases, Bath, UK
  • Volume
    2
  • fYear
    1999
  • fDate
    36434
  • Abstract
    Thermal infrared imaging (IR) in medicine has been dependent on single element detectors with mechanical scanning systems for four decades. Modern IR systems now offer high speed, high resolution. This paper reports a study made during an upgrade from a cooled single element Agena 782M imager to a new uncooled focal plane army Thermovision (Flir systems Inc.). The results show spatial resolution to be higher with the new Thermovision, resolving skin vasculature. The autocalibration of this camera can produce some temperature measurement variability
  • Keywords
    biomedical equipment; biomedical imaging; biothermics; blood vessels; cameras; focal planes; image resolution; skin; temperature measurement; Flir systems Inc.; camera autocalibration; clinical infra red imaging; cooled single element Agena 782M imager; focal plane array system; high speed high resolution systems; mechanical scanning systems; medical diagnostic imaging; single element detectors; spatial resolution; temperature measurement variability; Biomedical imaging; Calibration; Cameras; Detectors; Diseases; Image resolution; Optical imaging; Skin; Spatial resolution; Temperature measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.804282
  • Filename
    804282