    Haemodynamic trend detection using the median filter

  • Author

    Mäkivirta, A. ; Koski, E.

  • Author_Institution
    Tech. Res. Centre of Finland, Tampere, Finland
  • fYear
    1992
  • fDate
    17-19 Sep 1992
  • Firstpage
    9
  • Lastpage
    11
  • Abstract
    Attention is given to methods to obtain clinically useful trend data from invasively monitored haemodynamic parameters in an intensive care unit. The requirements for a trend detector in the context of intensive care are defined. A median filter-based trend detector designed to perform as a preprocessor for a highly specific knowledge-based alarm system is presented. The trend detector uses the standard median filter to preprocess the data before estimating first-order trends with least squares sum line fitting. The detector´s responses to artifact input and its reliability as well as its clinical evaluation are discussed
  • Keywords
    alarm systems; filtering and prediction theory; haemodynamics; knowledge based systems; parameter estimation; patient monitoring; first-order trends; haemodynamic trend detection; intensive care unit; knowledge-based alarm system; least squares sum line fitting; median filter-based trend detector; parameter estimation; patient monitoring; Blood flow; Change detection algorithms; Detectors; Heart rate; Least squares approximation; Low pass filters; Monitoring; Nonlinear filters; Predictive models; Testing;