DocumentCode
3320354
Title
Serials capacitors method of high capacitance calibration
Author
Wu Kang ; Liu Min ; Li Yalu
Author_Institution
Beijing Orient Inst. of Meas. & Test, Beijing, China
Volume
1
fYear
2013
fDate
16-19 Aug. 2013
Firstpage
464
Lastpage
467
Abstract
The paper proposes a new method for high capacitance calibration, which is called series capacitors method, including details of calibration principle, executive procedure and uncertainty analysis of the method for high capacitance from 10μF to 1mF. Because of the traceability and high resolution of the precision capacitance bridge AH2700A, the high capacitance can be calculated by the measurement of standard capacitor, series capacitor and distributed capacitance. The paper also describes the uncertainty analysis of the method, and gives the results of the uncertainty evaluation.
Keywords
calibration; capacitance measurement; capacitors; measurement uncertainty; capacitance 10 muF to 1 mF; distributed capacitance; executive procedure; high capacitance calibration; serials capacitors method; traceability; uncertainty analysis; Calibration; Capacitance; Capacitance measurement; Capacitors; Measurement uncertainty; Uncertainty; Wires; High capacitance; Series capacitors method; Traceability; Uncertainty analysis;