• DocumentCode
    3322107
  • Title

    Improved resolution and dynamic range in medical ultrasonic imaging using depth-dependent mismatched filtering

  • Author

    Brenner, A.R. ; Eck, K. ; Wilhelm, W. ; Noll, T.G.

  • Author_Institution
    Tech. Hochschule Aachen, Germany
  • Volume
    2
  • fYear
    1997
  • fDate
    5-8 Oct 1997
  • Firstpage
    1475
  • Abstract
    The authors present a depth-dependent mismatched-filter pulse compression technique which improves range resolution and dynamic range of coded excitation ultrasonic imaging compared to a matched-filter pulse compression. After choosing a suitable magnitude spectrum, frequency weighting is used to design the appropriate mismatched filter depth-dependently, which allows to control mainlobe width and sidelobe characteristics. Both simulations and measurements are used to demonstrate the efficacy of the technique. A VLSI implementation of a depth-dependent mismatched filter is shown to be feasible
  • Keywords
    VLSI; acoustic filters; biomedical ultrasonics; image resolution; medical image processing; pulse compression; VLSI implementation; coded excitation ultrasonic imaging; depth-dependent mismatched filtering; dynamic range improvement; frequency weighting; magnitude spectrum; mainlobe width; matched-filter pulse compression; medical diagnostic imaging; medical ultrasonic imaging; resolution improvement; sidelobe characteristics; Attenuation; Biomedical imaging; Chirp; Dynamic range; Frequency; Image resolution; Matched filters; Pulse compression methods; Signal resolution; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 1997. Proceedings., 1997 IEEE
  • Conference_Location
    Toronto, Ont.
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-4153-8
  • Type

    conf

  • DOI
    10.1109/ULTSYM.1997.661855
  • Filename
    661855