DocumentCode
332470
Title
X-band pulsed electron paramagnetic resonance spectrometer
Author
Duchiewicz, Jan ; Dobrucki, Andrzej L. ; Francik, Andrzej ; Czoch, Ryszard ; Sadowski, Andrzej ; Liber, Arkadiusz
Author_Institution
Inst. of Telecommun. & Acoust., Wroclaw Univ. of Technol., Poland
Volume
2
fYear
1998
fDate
20-22 May 1998
Firstpage
643
Abstract
X-Band pulsed EPR spectrometer designed and constructed in Wroclaw UT is described. The spectrometer has been designed as a modular system which makes it easy to upgrade. In the microwave path a power pulse amplifier with traveling wave tube (Pout=350 W) is used. Pulse EPR signals are recorded with a fast digital storage oscilloscope of LeCroy (9361, 2.5 GS/s), supported with hardware mathematical option (WP02) performing FFT calculations. The system is fully controlled with PC computer via a GPIB interface
Keywords
EPR spectrometers; digital storage oscilloscopes; fast Fourier transforms; microwave spectrometers; spectroscopy computing; FFT calculations; X-band; fast digital storage oscilloscope; hardware mathematical option; microwave path; modular system; power pulse amplifier; pulsed electron paramagnetic resonance spectrometer; Computer interfaces; Control systems; Electrons; Hardware; Microwave amplifiers; Oscilloscopes; Paramagnetic resonance; Power amplifiers; Pulse amplifiers; Spectroscopy;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwaves and Radar, 1998. MIKON '98., 12th International Conference on
Conference_Location
Krakow
Print_ISBN
83-906662-0-0
Type
conf
DOI
10.1109/MIKON.1998.740939
Filename
740939
Link To Document