DocumentCode :
3325132
Title :
A modified statistical system model and an accelerated Monte Carlo simulation for a silicon detector based Compton medical imaging system
Author :
Han, Li ; Clinthorne, Neal H.
Author_Institution :
Univ. of Michigan, Ann Arbor, MI, USA
fYear :
2009
fDate :
Oct. 24 2009-Nov. 1 2009
Firstpage :
2908
Lastpage :
2912
Abstract :
The objective of this research is to demonstrate a modified statistical system model and an approach of accelerated Monte Carlo simulation for studying the capability of a dual-planar Compton medical imaging system constructed with silicon and NaI detectors to image 364.4 keV energy photons emitted from I131. The Compton imaging system is a potential medical imaging instrument for effectively observing the process of radionuclide cancer treatments. Since decoupling the tradeoff between spatial resolution and detection efficiency inherent to the absorbing collimation of conventional Anger Camera, the Compton imaging system provides improved imaging performance in both detection efficiency and spatial resolution especially for higher energy photons. In the study, a modified statistical system model of the Compton system was developed considering all factors in the Compton process including Doppler broadening, energy resolution and spatial resolution of the scattering and absorbed detectors in addition to the Compton image formation process. To conquer the limitation of low simulation speed when employing the existent allpurpose Monte Carlo simulation platform due to low Compton interaction cross-section inside of the silicon detector, the developed Monte Carlo simulation involves the techniques of force detection and variance reduction to speed up the simulation. It is proved that two promotion approaches are of benefit to performance analysis and image reconstruction for the Compton imaging system.
Keywords :
Compton effect; Doppler broadening; Monte Carlo methods; cancer; image reconstruction; medical image processing; radioisotope imaging; silicon radiation detectors; solid scintillation detectors; statistical analysis; Compton image formation; Compton process; Doppler broadening; NaI detector; absorbing collimation; accelerated Monte Carlo simulation; conventional Anger Camera; dual-planar Compton medical imaging system; electron volt energy 364.4 keV; force detection; image reconstruction; low Compton interaction cross-section; modified statistical system model; radionuclide cancer treatments; silicon detector; variance reduction; Acceleration; Biomedical imaging; Cancer; Detectors; High-resolution imaging; Instruments; Medical treatment; Optical collimators; Silicon; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
ISSN :
1095-7863
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2009.5401616
Filename :
5401616
Link To Document :
بازگشت