DocumentCode
3326431
Title
Investigation of a sub-millimeter resolution PET detector with depth of interaction encoding using digital SiPM single sided readout
Author
Düppenbecker, Peter Michael ; Lodomez, Sarah ; Haagen, Rainald ; Marsden, Paul K. ; Schulz, Volkmar
Author_Institution
Philips Res. Eur., Aachen, Germany
fYear
2011
fDate
23-29 Oct. 2011
Firstpage
2252
Lastpage
2253
Abstract
One of the main challenges for building high resolution PET scanners with sub-millimeter resolution is to compensate for the parallax error induced by the height of the detector elements. During the last years several approaches to extract depth of interaction information for pixilated scintillation detectors were evaluated. We investigate the feasibility of a two layer PET detector with single sided readout using digital SiPM (dSiPM) arrays whereby the depth information is encoded in the width of the light distribution.
Keywords
nuclear electronics; positron emission tomography; readout electronics; scintillation counters; silicon radiation detectors; PET detector; depth-of-interaction encoding; depth-of-interaction information; detector elements; digital SiPM arrays; high resolution PET scanners; light distribution width; parallax error; pixilated scintillation detectors; single sided readout; sub-millimeter resolution; Photonics;