• DocumentCode
    3326431
  • Title

    Investigation of a sub-millimeter resolution PET detector with depth of interaction encoding using digital SiPM single sided readout

  • Author

    Düppenbecker, Peter Michael ; Lodomez, Sarah ; Haagen, Rainald ; Marsden, Paul K. ; Schulz, Volkmar

  • Author_Institution
    Philips Res. Eur., Aachen, Germany
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    2252
  • Lastpage
    2253
  • Abstract
    One of the main challenges for building high resolution PET scanners with sub-millimeter resolution is to compensate for the parallax error induced by the height of the detector elements. During the last years several approaches to extract depth of interaction information for pixilated scintillation detectors were evaluated. We investigate the feasibility of a two layer PET detector with single sided readout using digital SiPM (dSiPM) arrays whereby the depth information is encoded in the width of the light distribution.
  • Keywords
    nuclear electronics; positron emission tomography; readout electronics; scintillation counters; silicon radiation detectors; PET detector; depth-of-interaction encoding; depth-of-interaction information; detector elements; digital SiPM arrays; high resolution PET scanners; light distribution width; parallax error; pixilated scintillation detectors; single sided readout; sub-millimeter resolution; Photonics;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6152490
  • Filename
    6152490