Title :
Calibration of PET scanners with a new traceable point-like Ge68/Ga68 source
Author :
Hasegawa, Tomoyuki ; Oda, Keiichi ; Wada, Yasuhiro ; Sato, Yasushi ; Murayama, Hideo ; Yamada, Takahiro ; Matsumoto, Mikio ; Igarashi, Masafumi ; Iryo, Jun ; Kamitaka, Hiroaki
Author_Institution :
Allied Health Sci., Kitasato Univ., Sagamihara, Japan
Abstract :
We have been developing reliable and convenient calibration methods based on the use of point-like radioactive sources. In the present study, we used a newly developed traceable point-like Ge68/Ga68 radioactive source for determining calibration factors of a PET scanner. The point-like Ge68/Ga68 radioactive source consists of a spherical aluminum absorber with an outer diameter of 8 mm and a small radioactive bead in it. The physical characteristics of the source were evaluated with a Monte Carlo simulation code based on Geant4. The source radioactivity was calibrated at an accredited calibration facility. The point-like source was placed in the field-of-view of a clinical PET scanner, SET-2400W (Shimadzu Co.). The calibration factors determined with the point-like source were consistent with those determined by a conventional standard calibration method within a reasonable uncertainty. In conclusion, the proposed method with newly developed traceable point-like sources is useful for calibrating PET scanners.
Keywords :
Monte Carlo methods; calibration; gallium; germanium; positron emission tomography; radioactive sources; Ga; Ge; Ge68/Ga68 radioactive source; Geant4; Monte Carlo simulation code; PET scanner calibration; SET-2400W; clinical PET scanner; spherical aluminum absorber; traceable point-like radioactive sources; Calibration; Detectors; Positron emission tomography;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
Print_ISBN :
978-1-4673-0118-3
DOI :
10.1109/NSSMIC.2011.6152510