New attenuation correction for the HRRT using transmission scatter correction and total variation regularization
Author
Sibomana, Merence ; Keller, Sune Hogild ; Svarer, Claus ; Olesen, Oline Vinter ; Andersen, Flemming ; Holm, Soren ; Hjgaard, L.
fYear
2009
fDate
Oct. 24 2009-Nov. 1 2009
Firstpage
3284
Lastpage
3286
Abstract
In the standard software for the Siemens HRRT PET scanner the most commonly used segmentation in the ¿-map reconstruction for human brain scans is MAP-TR. Problems with bias in the lower cerebellum and pons in HRRT brain images have been reported. The main source of the problem is poor bone / soft tissue segmentation in these regions and the lack of scatter correction in the ¿-map reconstruction. In this paper we describe and validate the new TXTV segmentation method (included in the HRRTU 1.0 and 1.1 user software) aimed at solving the bias problem.
Keywords
brain; positron emission tomography; HRRT brain images; Siemens HRRT PET scanner; human brain scans; lower cerebellum; total variation regularization; transmission scatter correction; Attenuation; Biological tissues; Bones; Brain; Humans; Image reconstruction; Image segmentation; Positron emission tomography; Scattering; Software standards;