A sealed, UHV compatible, soft X-ray detector utilizing gas electron multipliers
Author :
Schaknowski, Neil A. ; Smith, Graham C. ; Yu, B. ; Fischer, Daniel A.
Author_Institution :
Instrum. Div., Brookhaven Nat. Lab., Upton, NY, USA
fYear :
2009
fDate :
Oct. 24 2009-Nov. 1 2009
Firstpage :
245
Lastpage :
249
Abstract :
An advanced soft X-ray detector has been designed and fabricated for use in synchrotron experiments that utilize X-ray absorption spectroscopy in the study a wide range of materials properties. Fluorescence X-rays, in particular CK at 277eV, are converted in a low pressure gas medium, and charge multiplication occurs in two gas electron multipliers, fabricated in-house from glass reinforced laminate, to enable single photon counting. The detector satisfies a number of demanding characteristics often required in synchrotron environments, such as UHV compatibility, compactness, long-term stability, and energy resolving capability.
Keywords :
X-ray fluorescence analysis; XANES; electron multiplier detectors; ionisation chambers; NEXAFS; X-ray absorption spectroscopy; fluorescence X-rays; gas electron multipliers; near-edge extended X-ray absorption fine structure; photon counting; soft X-ray detector; synchrotron environments; synchrotron experiments; Electromagnetic wave absorption; Electron multipliers; Fluorescence; Glass; Laminates; Material properties; Spectroscopy; Stability; Synchrotrons; X-ray detectors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
ISSN :
1095-7863
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2009.5401777
Filename :
5401777
Link To Document :
بازگشت