• DocumentCode
    3328327
  • Title

    Evaluation of APD and SiPM matrices as sensors for monolithic PET detector blocks

  • Author

    De Acilu, P. García ; Mendes, P. Rato ; Cañadas, M. ; Sarasola, I. ; Cuerdo, R. ; Romero, L. ; Willmott, C.

  • Author_Institution
    CIEMAT (Centro de Investig. Energeticas, Medioambientales y Tecnol.), Madrid, Spain
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    3300
  • Lastpage
    3304
  • Abstract
    Gamma detectors based on monolithic scintillator blocks coupled to APDs matrices have proved to be a good alternative to pixelated ones for PET scanners. They provide comparable spatial resolution, improve the sensitivity and make easier the mechanical design of the system. In this study we evaluate by means of Geant4-based simulations the possibility of replacing the APDs by SiPMs. Several commercial matrices of light sensors coupled to LYSO:Ce monolithic blocks have been simulated and compared. Regarding the spatial resolution and linearity of the detector, SiPMs with high photo detection efficiency could become an advantageous replacement for the APDs.
  • Keywords
    gamma-ray detection; positron emission tomography; silicon radiation detectors; APD matrix evaluation; Geant4-based simulations; PET scanners; SiPM matrix evaluation; comparable spatial resolution; gamma detectors; monolithic PET detector blocks; monolithic scintillator blocks; photo detection efficiency; system mechanical design; Sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6152594
  • Filename
    6152594