    Assessment of DOI resolution in a fully mounted PEM scanner

  • Author

    Matela, N. ; Ferreira, C.S. ; Martins, M.V. ; Bugalho, R. ; Neves, J.A. ; Ortigão, C. ; Oliveira, N. ; Varela, J. ; Almeida, P.

  • Author_Institution
    Inst. de Biofisica e Eng. Biomed., Univ. de Lisboa, Lisbon, Portugal
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    3351
  • Lastpage
    3353
  • Abstract
    The ClearPEM scanner is a dedicated system to perform breast imaging with the ability of measuring DOI, allowing an image spatial resolution close to 1.5 mm. The DOI is measured by calculating the amplitude asymmetry between the signals produced by two APD, one in the top and other in the bottom of the crystal. The DOI resolution measured in laboratory tests is 2.0 mm. In this paper we present a method to periodically confirm this value with the scanner fully mounted. The DOI resolution obtained with this method was 2.8 mm, which is consistent with the resolution measured during the laboratory tests.
  • Keywords
    avalanche photodiodes; biomedical equipment; mammography; positron emission tomography; solid scintillation detectors; APD; ClearPEM scanner; DOI resolution assessment; avalanche photodiodes; breast imaging; dedicated system; depth of interaction; fully mounted PEM scanner; image spatial resolution; positron emission tomography; signal amplitude asymmetry;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6152607
  • Filename
    6152607