  • Title

    MR compatible brain PET using tileable GAPD arrays

  • Author

    Jung, Jin Ho ; Choi, Yong ; Hong, Key Jo ; Kang, Ji Hoon ; Hu, Wei ; Min, Byung Jun ; Huh, Yoon Suk ; Shin, Seung Han ; Lim, Hyun Keong ; Kim, Dae Shik ; Jin, Han Byul

  • Author_Institution
    Sch. of Med., Dept. of Nucl. Med., Sungkyunkwan Univ., Seoul, South Korea
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    3556
  • Lastpage
    3559
  • Abstract
    The aim of this study is to develop a MR compatible PET that is insertable to MRI and allows simultaneous PET and MR imaging of human brain. The brain PET having 72 detector modules arranged in a ring of 330 mm diameter was designed. Each PET module composed of 4 ? 4 matrix of 3 mm ? 3 mm ? 20 mm LYSO crystals coupled to a tileable 4 ? 4 array Geigermode avalanche photodiode (GAPD) and designed to locate between RF and gradient coils. Signals of the each module were transferred to preamplifiers using flexible flat cable of 3 m long, and then sent to a position decoder circuit (PDC), which outputs digital address and an analog pulse of the one interacted channel from preamplifier signals. The PDC outputs were fed into FPGA-embedded DAQ boards. The analog signal was digitized, and arrival time and energy of the signal were calculated and stored. All electronics were located outside MR bore to minimize signal interference between PET and MR. Basic performance of the PET components and cross-compatibilities of the PET module and MR were evaluated. Imaging performance of the designed brain PET was investigated using Monte Carlo simulation and experimental measurement. The degradation of PET performance caused by the 3 m long cable and the PDC was negligibly small. No obvious differences of the PET module performance measured inside/outside MR bore were observed. The SNR of various MR sequence phantom images acquired with/without the PET module were also similar. Activity distribution patterns of hot-rod phantoms were well imaged without distortion, and rods down to a diameter of 3.5 mm were resolved in both simulation and experiment. Grayand white matter of the Hoffman brain phantom was also well imaged. Preliminary experimental results demonstrate that MR compatible high quality PET imaging is feasible using the GAPD arrays, electronics, signal processing method and MR insertable PET design schemes developed in this study.
  • Keywords
    Geiger counters; Monte Carlo methods; avalanche photodiodes; biomedical MRI; brain; image sequences; medical image processing; phantoms; positron emission tomography; FPGA; GAPD arrays; Geigermode avalanche photodiode; LYSO crystals; MRI; Monte Carlo simulation; PET; RF coils; brain; gradient coils; gray matter; image sequence; phantom; position decoder circuit; white matter; Boring; Coupling circuits; Crystals; Detectors; Humans; Imaging phantoms; Magnetic resonance imaging; Positron emission tomography; Power cables; Preamplifiers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5401815
  • Filename
    5401815