Slit-slat collimator geometrical calibration for a PET/SPECT dual modality animal scanner
Author :
Deng, Xiao ; Beaudoin, Jean-Francois ; Cadorette, Jules ; Ma, Tianyu ; Lecomte, Roger ; Yao, Rutao
Author_Institution :
Dept. of Nucl. Med., State Univ. of New York at Buffalo, Buffalo, NY, USA
Abstract :
As part of our continued effort on developing a dual modality PET/SPECT imaging device based on an animal PET scanner, we studied two new techniques for calibrating the geometrical parameters of the slit-slat collimator insert based SPECT system. The collimator´s slat-to-crystal position (SCP) in axial direction was determined from the cyclic pattern of the sensitivity profile of a series of 99mTc point source measurements in the axial direction. To calibrate the center of rotation (COR) and slit aperture centers (SACs), a 18F line source was first attached along one of the slits and imaged in PET mode at multiple rotational positions of the slit. The COR and SAC values were then estimated from the PET images of the line source. The calibration techniques were assessed with Monte Carlo simulated and experimental studies. The calibrated geometrical parameters were used in analytical calculation to generate the system matrix for SPECT image reconstruction. Phantom and animal studies were performed to demonstrate the imaging performance. The proposed new calibration methods were found effective in enabling slit-slat SPECT imaging on a PET scanner.
Keywords :
Monte Carlo methods; calibration; collimators; fluorine; image reconstruction; image scanners; medical image processing; positron emission tomography; sensitivity; single photon emission computed tomography; technetium; 18F line source; 99mTc point source measurements; Monte Carlo simulation; PET-SPECT dual modality animal scanner; axial direction; cyclic pattern; image reconstruction; multiple rotational positions; phantom; sensitivity profile; slit aperture centers; slit-slat collimator geometrical calibration; Biomedical imaging; ISO; ISO standards; Logic gates; Positron emission tomography;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
Print_ISBN :
978-1-4673-0118-3
DOI :
10.1109/NSSMIC.2011.6152663