Title :
Adaptive normalized metal artifact reduction (ANMAR) in computed tomography
Author :
Meyer, Esther ; Raupach, Rainer ; Schmidt, Bernhard ; Mahnken, Andreas H. ; Kachelriess, Marc
Author_Institution :
Institute of Medical Physics (IMP), University of Erlangen-Nürnberg, Henkestr. 91, 91052, Germany
Abstract :
Metal artifacts drastically impair the image quality in CT images and often reduce their diagnostic value. There are many publications on metal artifact reduction (MAR) which regard the metal-affected parts of the rawdata as completely unreliable and therefore replace it. While those sinogram inpainting methods are in general successful in removing metal artifacts, they cannot exactly recover the true values which are replaced and therefore the corrected image will exhibit new artifacts and blurring.
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
Print_ISBN :
978-1-4673-0118-3
DOI :
10.1109/NSSMIC.2011.6152691