DocumentCode
3331152
Title
The Improvement of Spectrometric Determination of COD by Microwave Digestion with Digestion Vessels
Author
Zhuang, Wen ; Luo, Jianzhong
Author_Institution
Fac. of Environ. Sci. & Eng., Guangdong Univ. of Technol., Guangzhou, China
fYear
2011
fDate
10-12 May 2011
Firstpage
1
Lastpage
4
Abstract
The pressure tightness of digestion vessels and the stability of voltage had effects on the spectrometric determination of COD by microwave digestion with digestion vessels. After the samples were digested, it was found that the each digestion vessels has different degree of water evaporation. Given everything equals except the voltage stability, there were biggish changes in the absorbance of blank samples after different batches of samples were digested. In order to eliminate the effect of water evaporation, the samples were poured out and calibrated to the same scale for spectrometric determination after being digested and cooled. The result showed that the measured values of COD were more accurate after the improvement and the correlation rate of the standard curve reached to 0.9999. Additionally, adding a voltage stabilizer to ensure the stable voltage during the process of microwave digestion was suggested.
Keywords
evaporation; water pollution; chemical oxygen demand; digestion vessels; microwave digestion; pressure tightness; spectrometric determination; voltage stability; voltage stabilizer; water evaporation; water pollution; Chemicals; Electromagnetic heating; Microwave devices; Microwave measurements; Microwave oscillators;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioinformatics and Biomedical Engineering, (iCBBE) 2011 5th International Conference on
Conference_Location
Wuhan
ISSN
2151-7614
Print_ISBN
978-1-4244-5088-6
Type
conf
DOI
10.1109/icbbe.2011.5780738
Filename
5780738
Link To Document