Title :
SiPM based detector module and digital data acquisition system for PET: Initial results
Author :
Safavi-Naeini, M. ; Franklin, D.R. ; Petasecca, M. ; Lerch, M.L.F. ; Kirkham, R. ; Moorhead, G. ; Dunn, P. ; De Geronimo, G. ; Rosenfeld, A.B.
Author_Institution :
Centre for Med. Radiat. Phys., Univ. of Wollongong, Wollongong, NSW, Australia
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
A novel detector module configuration based on silicon photomultiplier (SiPM) detectors with LYSO scintillators, a multichannel mixed analog/digital ASIC and an FPGA-based digital data acquisition system has been developed for preclinical positron emission tomography (PET). The SCEPTER ASIC, originally developed for soft X-ray detection at Brookhaven National Laboratory, has a unique architecture incorporating multiple peak detectors and time-to-amplitude converters, and offers an extremely low dead time, making it ideal for coincidence detection. A representative pair of channels in the prototype system has been characterised. Measured energy and timing resolutions are reported for coincidences between a single-pixel SiPM detector and a conventional PMT. A timing resolution of 11.7 ns was achieved during a 48 hour run. Energy resolution for the SiPM device and data acquisition system was found to be 17%.
Keywords :
biomedical equipment; coincidence techniques; field programmable gate arrays; mixed analogue-digital integrated circuits; photomultipliers; positron emission tomography; scintillation counters; semiconductor devices; FPGA based digital data acquisition; LYSO scintillators; PET; SCEPTER ASIC; SiPM based detector module; application specific integrated circuit; coincidence detection; conventional PMT; detector module configuration; digital data acquisition system; energy resolution; field programmable gate array; multichannel mixed analog-digital ASIC; peak detectors; positron emission tomography; silicon photomultiplier detectors; single pixel SiPM detector; time to amplitude converters; timing resolution; Application specific integrated circuits; Data acquisition; Energy resolution; Laboratories; Photomultipliers; Positron emission tomography; Silicon; Timing; X-ray detection; X-ray detectors;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2009.5401989